| Measurement Method | 
Ratio turbidimetric determination using a primary nephelometric light | 
| scatter signal (90°) to the transmitted light scatter signal | 
| Regulatory | 
Meets EPA Method 180.1 | 
| Lamp Source | 
Tungsten filament lamp | 
| Range | 
0-1000 NTU | 
| Accuracy | 
±2% of reading plus stray light from 0-1000 NTU | 
| Repeatability | 
±1% of reading or 0.01 NTU, whichever is greater | 
| Resolution | 
0.01 NTU on lowest range | 
| Stray Light | 
= 0.02 NTU | 
| Signal Averaging | 
Selectable on or off | 
| Detector | 
Silicon Photodiode | 
| Reading Modes | 
Normal (Push to Read) | 
| Signal Averaging  | 
| Rapidly Settling Turbidity(TM) | 
| Calibration Options | 
Single step RapidCal(TM) for low-level regulatory reporting from 0-40 NTU | 
| Full range calibration from 0-1000 NTU | 
| Calibration to degrees of turbidity | 
| Calibration Logger | 
Records the last 25 successful calibrations | 
| Verification Logger | 
Logs the last 250 successful verifications | 
| Data Logger | 
500 records | 
| Power Requirement | 
AC 100-240 V, 50/60 Hz (with Power or USB+Power Module) | 
| 4 AA alkaline batteries | 
| Rechargeable NiMH (with USB+Power Module) | 
| Operating Conditions | 
Temperature:  0 to 50 °C (32 to 122 °F) | 
| Relative Humidity:  0-90% at 30 °C, 0-80% at 40 °C, 0-70% at 50 °C noncondensing | 
| Storage Conditions | 
-40 to 60 °C (-40 to 140 °F), instrument only | 
| Interface | 
Optional USB | 
| Sample Required | 
15 mL (0.5 oz.) | 
| Sample Cells | 
Round cells, 60 x 25 mm (2.36 x 1 in.), borosilicate glass with screw caps | 
| Dimensions | 
22.9 x 10.7 x 7.7 cm (9.0 x 4.2 x 3.0 in.) | 
| Weight | 
530 g (1.17 lb) without batteries | 
| 620 g (1.37 lb) with four AA alkaline batteries | 
| Meter Enclosure Rating | 
IP67 (closed lid, battery and module compartment excluded) | 
| Protection Class | 
Power supply: Class II | 
| Certification | 
CE certified | 
| Warranty | 
1 year |